Mainstream vs Sidestream Capnography: How to Source a Compatible EtCO2 Sensor
How mainstream and sidestream capnography differ, and how to source and verify a compatible EtCO2 sensor for your patient monitor. A buyer's guide.
Read articleMedten-distributed fetal replacement parts for Philips Avalon FM20, FM30, FM40, and FM50 are designed, material-selected, and IP68-tested to match the water-ingress protection designated for this equipment — including 12-hour immersion at 1.1 m, stricter than the OEM baseline of 1 m for 5 hours.

Medten-distributed fetal replacement parts for Philips Avalon FM20, FM30, FM40, and FM50 match OEM IP68 water-ingress protection through four evidence streams: a four-lip star (X-ring) seal, disinfection-resistant elastomer, groove dimensions matched to the OEM housing, and 100% CCD inspection of every seal lot. Qualification immerses powered-off assembled transducers in 1.1 m of water for 12 hours — stricter than the OEM baseline of 1 m for 5 hours — then opens them to confirm zero water ingress and full function. These are repair-only service parts for qualified biomedical technicians, not OEM devices and not for end-user self-install.
Who this guide is for: biomedical engineers, clinical engineers, and service technicians repairing Philips Avalon FM20, FM30, FM40, and FM50 fetal transducers — and procurement teams qualifying replacement parts for a service line. It is a repair-parts and ingress-protection guide, not clinical fetal-monitoring guidance.
When a Philips Avalon fetal monitor transducer fails in service — a cracked housing, a swollen seal, a leak after cleaning — biomedical engineers and service technicians need a replacement part they can trust to keep water and fluids out for the remaining life of the device. This article explains how Orantech (Medten’s manufacturing partner) replacement parts for use with Avalon FM20, FM30, FM40, and FM50 are designed, material-selected, and IP68-tested to deliver water-ingress protection that matches the IP68 standard designated for this equipment.
IP68 is the ingress-protection code defined in the international standard IEC 60529. The two digits describe the enclosure’s resistance to solid particles (6 = dust-tight) and water (8 = continuous immersion beyond 1 m, as specified by the manufacturer). For a bedside transducer that is cleaned with wipes, gels, and occasional fluids, this rating is not a marketing line — it is the difference between a reliable FHR/Toco trace and an electronics failure.
Water ingress is the leading end-of-life event for fetal transducers. Once it reaches the coupling membrane or the PCB, signal quality degrades long before the device fails outright. A replacement part that cannot hold IP68 in the field simply shortens the service interval the OEM designed for.
Replacement service parts are not intended to be a shortcut. Our non-OEM replacement transducers and seal assemblies for use with Avalon FM20/30/40/50 are engineered to provide IP68-equivalent ingress protection through four parallel evidence streams:
The star (or “X”) cross-section creates four independent lips that contact the mating surface simultaneously. If one lip wears or is contaminated, three remain. This is the same redundancy principle used in higher-pressure fluid-seal design and is what lets the seal keep working after a technician opens the housing for cleaning or service.
Hospital cleaning protocols put elastomers through a punishing cycle: alcohol wipes, quaternary-ammonium disinfectants, gel residue, and skin oils. We select seal compounds on four numbers: abrasion resistance, low coefficient of friction (so the housing slides without rolling the seal), operating-temperature range, and aging resistance.
A seal only works if the groove holds it at the right compression. We match the OEM housing groove, which gives the four-lip geometry room to do its job.
IP codes are pass/fail, but the method behind the pass is what matters. Here is what we run on every qualification lot, and what the original-equipment specification indicates by comparison.
| Test | OEM protection baseline | Medten replacement-part qualification |
|---|---|---|
| Dust ingress (IP6X) | Dust shall not penetrate the enclosure; signal acquisition unaffected. | Assembled transducer placed in a dust chamber; 8 hours of negative-pressure dust cycling; unit opened and inspected internally for zero dust ingress and full function. |
| Water ingress (IPX8) | 1 m depth, 5 hours continuous immersion; no ingress; full function on retrieval. | Powered-off unit fully submerged in 1.1 m of water for 12 hours; unit opened and inspected internally for zero water ingress and full function on retrieval. |
Six transducers go in. They come out wet. They are opened and inspected by hand. Every cycle ends in a pass/fail on the same checklist: zero water, full function.
A repair bay sees mixed inventory. To confirm our seal works regardless of which housing stack it lands in, we run a six-cell mix-and-match matrix that crosses OEM and Medten upper shells, lower shells, and seal rings. Every cell passes 12 hours of immersion.
| Combination | Top case | Bottom case | Seal ring | Quantity | Immersion time | Test result |
|---|---|---|---|---|---|---|
| 1 | Orantech | OEM | OEM | 2 | 12 hours | Pass |
| 2 | Orantech | OEM | Orantech | 2 | 12 hours | Pass |
| 3 | Orantech | Orantech | Orantech | 2 | 12 hours | Pass |
| 4 | OEM | OEM | Orantech | 2 | 12 hours | Pass |
| 5 | OEM | Orantech | OEM | 2 | 12 hours | Pass |
| 6 | OEM | Orantech | Orantech | 2 | 12 hours | Pass |
Each original part number below links to a Medten search that groups the repair parts we supply for servicing that transducer — seal kits, X-rings, and the other components a biomedical technician may need during a repair. If you do not see your part number on the list, contact us — we are adding coverage continuously.
| Original transducer (Philips) | Function | Medten-distributed replacement parts |
|---|---|---|
| 867245 | Toco + MP Transducer | Browse repair parts |
| 867249 | Toco + Transducer | Browse repair parts |
| 867248 | Toco MP Transducer | Browse repair parts |
| 867246 | Ultrasound Transducer | Browse repair parts |
| 867247 | ECG/IUP Transducer | Browse repair parts |
| 866075 | CL Toco+MP Transducer | Browse repair parts |
| 866076 | CL Ultrasound Transducer | Browse repair parts |
| 866077 | CL ECG/IUP Transducer | Browse repair parts |
Repair-only use. Medten-Distributed replacement parts are designed for installation by qualified biomedical service technicians during repair of the device. They are not OEM devices and are not offered as new medical-device replacements to end users. Always follow the original equipment manufacturer’s service documentation.
If you are evaluating replacement transducers for an Avalon FM20/30/40/50 service line, or you have a part number not on the cross-reference table, our engineering team can review your application, share the test data behind any specific lot, and quote lead times.
Contact Medten repair-parts engineering
Trademark notice. Philips, Avalon, FM20, FM30, FM40 and FM50 are registered trademarks of their respective owners; Medten USA Inc. is not affiliated with, sponsored by, licensed by, or endorsed by Philips, and such affiliation, sponsorship, licensure, or endorsement is in no way expressed or implied by Medten USA Inc. Use of any trademarks is to demonstrate compatibility or for comparison purposes only.
Qualified personnel notice. All Medten-Distributed replacement parts are service parts for use by qualified biomedical technicians.
IP rating notice. IP ratings refer to enclosure ingress protection per IEC 60529 and do not constitute sterilization or infection-control claims.
Keep reading
More articles you may find helpful from the Medten blog.